Single-chamber AAI mode

Tracing
N° 4
Manufacturer Medtronic Device PM Field Pacing Modes
Patient

Same patient as in tracing 1.

Graph and trace

Programming in AAI mode 60 beats/minute;

  1. atrial pacing at the base rate (AP) at the end of the escape interval (1000 ms between 2 atrial pacings);
  2. proper sensing of the atrial extrasystole (AS) and inhibition of atrial pacing;
  3. blocked atrial extrasystole properly sensed but not followed by ventricular pacing.
Comments

A single-chamber pacemaker operates in AAI mode when only one lead is positioned in the atrium; the AAI mode can also be programmed in a dual-chamber pacemaker. The AAI mode provides single-chamber atrial pacing at the programmed pacing rate unless inhibited by a sensed event. Sensing only applies to the atrium. The programming and functioning are essentially the same as for a VVI pacemaker. However, a higher sensitivity must be programmed (lower programmed value), since the amplitude of the atrial complexes is often lower than that of the ventricular complexes. Similarly, the refractory period must be longer so as to avoid ventricular oversensing. The detection of an R wave by an AAI pacemaker leads to a decrease in pacing rate, the R wave recycling the escape interval. In the presence of far-field R wave oversensing, it is possible to render the pacemaker less sensitive and/or prolong the refractory period.

This tracing highlights the main characteristics of this pacing mode:

  1. sensing and pacing of the atrium, inhibition due to a spontaneous event; the preferred indication is therefore pure sinus dysfunction without ventricular conduction disorder. An AAI pacemaker allows limiting the number of implanted leads, ensures a physiological rate at rest and during exercise after programming of the rate responsive frequency, and avoids unnecessary ventricular pacing;
  2. absence of ventricular sensing or pacing. AAI single-chamber pacemakers or AAI mode programming in a dual-chamber pacemaker are formally contraindicated in patients with a permanent or paroxysmal atrioventricular conduction disorder. They should also be avoided in patients with vagal symptoms or carotid sinus syndrome. This tracing shows the ventricular pauses appearing during the occurrence of a blocked atrioventricular extrasystole.

A recent study reported unfavorable results in the implantation of a single-chamber AAI pacemaker compared to the implantation of a dual-chamber in patients with sinus dysfunction. Indeed, the re-intervention rate was greater in patients implanted with an AAI, with the need to add a ventricular lead secondarily to the occurrence of an atrioventricular conduction disorder. More surprisingly, the incidence of atrial fibrillation episodes was also higher. The indications for implantation of a single-chamber AAI pacemaker are thus nowadays relatively limited.   

Other articles that may be of interest to you
Atrial pacing failure
EGM recordings