Optimization of the VV delay

Tracing
N° 8
Manufacturer Boston Scientific Device CRT Field AV & VV delays optimization
Patient

This 67-year-old man (same patient as tracing 3) received a Boston Scientific Cognis triple chamber defibrillator in the context of a dilated cardiomyopathy with left bundle branch block. The VV delay was re-programmed and tracings were recorded during an ambulatory visit.

Graph and trace
  1. RV pacing.
  2. BiV stimulation with RV pre-activation 40 ms ahead of LV stimulation.
  3. Simultaneous BiV stimulation.
  4. BiV stimulation with LV pre-activation 40 ms ahead of RV stimulation.
  5. LV stimulation.
Comments

These tracings show the effects of changes in the VV delay on the QRS morphology, including a change in axis between RV and LV stimulation. The morphology of the QRS is a rough indicator of the ventricular activation sequence, which is the target of BiV resynchronization. The optimal configuration should produce a ventricular activation associated with an optimal long-term clinical response. Combining the various VV and AV delays (in this patient, the setting of the AV delay produced variable amounts of fusion), the short-term programming choices are innumerable, though no single parameter currently predicts accurately an optimal long-term clinical response. The automatic settings directly chosen by the device have the advantage of a certain degree of reproducibility, at rest as well as during exercise, though their pathophysiologic rationale and clinical merit remain to be confirmed.

Other articles that may be of interest to you
atrial sensing failure
EGM recordings