ventricular AutoCaptureTM

Tracing
N° 16
Manufacturer Abbott Device PM
Graph and trace

Threshold measurement performed manually in the same patient; effective capture at 0.75 V and loss of capture at 0.5 V confirming the validity of the automatic threshold measurement.