Quizz - The 1st World Championship of device EGMs interpretation

Question

Your answerChoiceFeedbackCorrect answer
P wave oversensing
T wave oversensing
VT
RV lead fracture/insulation breach 

GOOD ANSWER : RV lead fracture/insulation breach

•Cloud of points on the tachogram. No train tracks. Atrial rate remains stable.

 
 
 
 
Should have chosen
Selected
AF
Question

Your answerChoiceFeedbackCorrect answer
P wave oversensing
T wave oversensing
EMI (electromagnetic interference)
Lead fracture
Selected
VF/VT

GOOD ANSWER : VF/VT 

Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
AF
Selected
T wave oversensing

GOOD ANSWER : T wave oversensing

 

 

T-Wave Oversensing: “Train tracks”

Why do the train tracks begin to separate?

•Notice that, in general, the fast (bottom) part of the railroad tracks remains steadily at 300ms.
•This represents a consistent interval between the R and T waves during the episode.
T-Wave Oversensing: “Train tracks”
•Medtronic interval plots are not completely linear. Notice each “notch” near the 0-axis represents 100ms, but toward the top of the graph each notch represents 300ms.
•Therefore the atrial rate, which appears to be consistent at about 600ms, is actually slowing after each shock. The ventricular rate, therefore, slows as well. 
 
Should have chosen
RV Lead Dislodgement
Lead fracture
VF
Question

Your answerChoiceFeedbackCorrect answer
Selected
RV Lead Dislodgement/ P wave oversensing

GOOD ANSWER : RV Lead Dislodgement/ P wave oversensing

•RV Lead Dislodgement: One month post device implant. Double counting. Double counting of P wave and R wave. Look at the interval between oversensed signals and most consistent with PR interval. Look at morphology of oversensed signal as well and most consistent with far field P wave.

 
Should have chosen
AF
EMI (electromagnetic interference)
Lead fracture
VF
Question

Your answerChoiceFeedbackCorrect answer
Ventricular tachycardia
Dual tachycardia (VT+ SVT)
Selected
SVT
SVT + aberrant atrioventricular conduction

GOOD ANSWER : SVT + aberrant atrioventricular conduction 

RID - : morphology different from the reference; 3 possibilities: VT, dual tachycardia, SVT + aberrant conduction
Afib: 2 possibilities: dual tachycardia, SVT + aberrant conduction
Unstb: in favor of SVT + aberrant conduction  

Should have chosen
Dual tachycardia (sinus tachycardia + VT)
Question

Your answerChoiceFeedbackCorrect answer
Sinus tachycardia
Selected
Slow VT

GOOD ANSWERS :

  • Slow VT
  • R-wave double counting

 

Slow VT episode with AV dissociation. Very wide ventricular EGMs were double sensed. The second ventricular EGM was sensed after the post-sensed ventricular refractory period. 

Should have chosen
R-wave double counting

GOOD ANSWERS :

  • Slow VT
  • R-wave double counting

 

 

Slow VT episode with AV dissociation. Very wide ventricular EGMs were double sensed. The second ventricular EGM was sensed after the post-sensed ventricular refractory period. 

Should have chosen
P-wave oversensing
T-wave oversensing
Question

Your answerChoiceFeedbackCorrect answer
Polymorphic VT
Ventricular undersensing
R-wave double counting
P-wave oversensing
Selected
T-wave oversensing

GOOD ANSWER : T-wave oversensing

 

•T-wave oversensing

•Oversensing of the T wave following a spontaneous ventricular event occurs preferentially in presence of a low-amplitude R wave. Sensitivity is are adjusted automatically, based on the amplitude of the R wave that was just sensed: when that amplitude is low, the likelihood is high of subsequently and rapidly reaching the highest sensitivity. Oversensing is facilitated by exercise, since exercise may be associated with a decrease in the R wave and an increase in T wave amplitude. The episode that was detected fell in the VT monitor zone, avoiding the delivery of any inappropriate therapy. 
Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
This is a burst during the charge  

GOOD ANSWER : This is a burst during the charge  

Fast tachycardia detected in the VF zone; diagnosis of VF after 12 F cycles;

burst during the charge + capacitors charge;
termination of the arrhythmia; however, the number of VS was insufficient to fill the counter of sinus rhythm restoration (only 3 VS cycles instead of 5 needed: return to sinus programmed to 5); continuation of the capacitors charge;
early recurrence of the arrhythmia;
electrical shock at the end of charge;
interruption of the arrhythmia and diagnosis of return to sinus rhythm.
 

Should have chosen
Selected
This is a burst before the charge
The burst is inefficient in reducing the VT
Selected
The parameter « return to sinus » is programmed to 3
After the ATP the first ventricular signal (-) is undersensed
Question

Your answerChoiceFeedbackCorrect answer
The programmed pacing mode is DDI
The programmed minimal pacing rate is 45 bpm
Selected
The programmed minimal pacing rate is 60 bpm

GOOD ANSWERS :

  • The programmed minimal pacing rate is 60 bpm
  • The maximal duration for an atrial pause is 1 second

 

The programmed pacing mode is DDD with Rythmiq On (AAI with VVI backup pacing); the minimal programmed pacing rate is 60 bpm since VVI pacing is delivered at 45 bpm (15 bpm slower than the programmed minimal pacing rate); therefore the maximal duration of the ventricular pause is 1333 ms (45 bpm) and the maximal duration of the atrial pause is 1OOO ms (60 bpm) 

Should have chosen
Selected
The maximal duration for a ventricular pause is 1.5 second
The maximal duration for an atrial pause is 1 second

GOOD ANSWERS :

  • The programmed minimal pacing rate is 60 bpm
  • The maximal duration for an atrial pause is 1 second

 

The programmed pacing mode is DDD with Rythmiq On (AAI with VVI backup pacing); the minimal programmed pacing rate is 60 bpm since VVI pacing is delivered at 45 bpm (15 bpm slower than the programmed minimal pacing rate); therefore the maximal duration of the ventricular pause is 1333 ms (45 bpm) and the maximal duration of the atrial pause is 1OOO ms (60 bpm) 

Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
Marker 1: Wenckebach behaviour and blocked P-wave
Marker 1: inappropriate intervention of the anti-PMT algorithm and blocked P-wave

GOOD ANSWERS :

  • Marker 1: inappropriate intervention of the anti-PMT algorithm and blocked P-wave
  • Intervention Rythmiq algorithm (Search AV +) and blocked P-waves

 

Marker 1: sinus tachycardia; after 16 consecutive cycles AS-VP-MT (spontaneous atrial activation and paced ventricle at the maximal tracking rate) with stable VP-AS interval, suspicion of PMT and prolongation of the PRAPV during one cycle explaining the cycle (AS) and the absence of ventricular pacing on one cycle

 

Marker 2: sustained aspect of 2/1 corresponding to the intervention of the Rythmiq algorithm (search AV +); once the device has switched to DDD mode, prolongation of the AV delay during 8 consecutive cycles to search for a spontaneous ventricular activation; in this patient during exercise, the prolongation of the AV delay induces this aspect of 2/1 (1/2 atrial activation is in the PRAPV);

Should have chosen
Marker 2: Wenckebach behavior and blocked P-waves
Marker 2:intervention of the anti-PMT algorithm and blocked P-waves
Selected
Intervention Rythmiq algorithm (Search AV +) and blocked P-waves

GOOD ANSWERS :

  • Marker 1: inappropriate intervention of the anti-PMT algorithm and blocked P-wave
  • Intervention Rythmiq algorithm (Search AV +) and blocked P-waves

 

Marker 1: sinus tachycardia; after 16 consecutive cycles AS-VP-MT (spontaneous atrial activation and paced ventricle at the maximal tracking rate) with stable VP-AS interval, suspicion of PMT and prolongation of the PRAPV during one cycle explaining the cycle (AS) and the absence of ventricular pacing on one cycle

 

Marker 2: sustained aspect of 2/1 corresponding to the intervention of the Rythmiq algorithm (search AV +); once the device has switched to DDD mode, prolongation of the AV delay during 8 consecutive cycles to search for a spontaneous ventricular activation; in this patient during exercise, the prolongation of the AV delay induces this aspect of 2/1 (1/2 atrial activation is in the PRAPV);

Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
Selected
Biotronik: 8

GOOD ANSWERS  :

  • Biotronik: 8
  • Boston Scientific: 16
  • SJM: 8

In Boston Scientific devices, the number of required consecutive VP-AS cycles is 16
For the other 4 companies, 8 consecutive VP-AS are required

Should have chosen
Selected
Boston Scientific: 16

GOOD ANSWERS  :

  • Biotronik: 8
  • Boston Scientific: 16
  • SJM: 8

 

In Boston Scientific devices, the number of required consecutive VP-AS cycles is 16
For the other 4 companies, 8 consecutive VP-AS are required

Should have chosen
Medtronic: 12
SJM: 8

GOOD ANSWERS  :

  • Biotronik: 8
  • Boston Scientific: 16
  • SJM: 8

 

In Boston Scientific devices, the number of required consecutive VP-AS cycles is 16
For the other 4 companies, 8 consecutive VP-AS are required

Should have chosen
Sorin: 6
Question

Your answerChoiceFeedbackCorrect answer
Appropriate Rythmiq commutation with second degree AV block
Appropriate Rythmiq commutation with third degree AV block
Selected
Ventricular undersensing

GOOD ANSWERS : 

  • Ventricular undersensing
  • Inappropriate Rythmiq-commutation to DDD (no AV block)

Inappropriate commutation to DDD-mode (no AV-block) induced by ventricular undersensing

Should have chosen
Selected
Inappropriate Rythmiq-commutation to DDD (no AV block)

GOOD ANSWERS : 

  • Ventricular undersensing
  • Inappropriate Rythmiq-commutation to DDD (no AV block)

Inappropriate commutation to DDD-mode (no AV-block) induced by ventricular undersensing

Should have chosen
Far-field P-wave oversensing
Question

Your answerChoiceFeedbackCorrect answer
LV pacing threshold: 2.5 V
Selected
LV pacing threshold: 2.75 V
Inefficient LV pacing

GOOD ANSWERS :

  • Inefficient LV pacing
  • Atrial capture with the LV pacing lead

Ineffective LV pacing; LV pacing captures the left atrium; atrial depolarization is sensed by the right atrial channel after a conduction delay between right and left atria;
The LV lead had dislodged.

Should have chosen
Selected
Atrial capture with the LV pacing lead

GOOD ANSWERS :

  • Inefficient LV pacing
  • Atrial capture with the LV pacing lead

Ineffective LV pacing; LV pacing captures the left atrium; atrial depolarization is sensed by the right atrial channel after a conduction delay between right and left atria;
The LV lead had dislodged.

Should have chosen
Right ventricular capture with the LV pacing lead
Question

Your answerChoiceFeedbackCorrect answer
Selected
Atrial tachycardia

GOOD ANSWERS :

  • Atrial tachycardia
  • Oversensing of the post BIV pacing signal with the atrial lead
  • Oversensing of the atrial activation with the RV lead

Atrial tachycardia and BiV stimulation. Paced ventricular events are sensed in the atrium (V/A cross-talk).
A/V oversensing with sensing of the atrial signal by the ventricular channel and inhibition of BiV stimulation, resulting in a 2-3 seconds pause. 

Should have chosen
Oversensing of the post BIV pacing signal with the atrial lead

GOOD ANSWERS :

  • Atrial tachycardia
  • Oversensing of the post BIV pacing signal with the atrial lead
  • Oversensing of the atrial activation with the RV lead

Atrial tachycardia and BiV stimulation. Paced ventricular events are sensed in the atrium (V/A cross-talk).
A/V oversensing with sensing of the atrial signal by the ventricular channel and inhibition of BiV stimulation, resulting in a 2-3 seconds pause. 

Should have chosen
Selected
Ventricular tachycardia
Conducted atrial tachycardia
Oversensing of the atrial activation with the RV lead

GOOD ANSWERS :

  • Atrial tachycardia
  • Oversensing of the post BIV pacing signal with the atrial lead
  • Oversensing of the atrial activation with the RV lead

 

Atrial tachycardia and BiV stimulation. Paced ventricular events are sensed in the atrium (V/A cross-talk).
A/V oversensing with sensing of the atrial signal by the ventricular channel and inhibition of BiV stimulation, resulting in a 2-3 seconds pause. 

Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
For the device, this is an episode of Ventricular Tachycardia
Selected
For the device, this is an episode of SVT

GOOD ANSWERS : 

  • For the device, this is an episode of SVT
  • The discrimination of this episode is only based on the analysis of the onset parameter

 

This is an episode of VT diagnosed SVT by the device based on single-lead discrimination: for the device the absence of onset is in favor of a sinus tachycardia. 

Should have chosen
The diagnosis of VT is based on a discrimination based on the PR Logic
The diagnosis of SVT is based on a discrimination based on the PR Logic
Selected
The discrimination of this episode is only based on the analysis of the onset parameter

GOOD ANSWERS : 

  • For the device, this is an episode of SVT
  • The discrimination of this episode is only based on the analysis of the onset parameter

 

This is an episode of VT diagnosed SVT by the device based on single-lead discrimination: for the device the absence of onset is in favor of a sinus tachycardia. 

Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
Cross-talk VA
Cross-talk AV

GOOD ANSWERS :

  • Cross-talk AV
  • T-wave oversensing
  • In Biotronik pacemakers, the duration of the post-atrial pacing ventricular safety window is 100 ms

 

In this tracing we observe post-VP and post-VS T wave oversensing and AV crosstalk with pacing at the end of the safety window (100 ms)

Should have chosen
Selected
T-wave oversensing

GOOD ANSWERS :

  • Cross-talk AV
  • T-wave oversensing
  • In Biotronik pacemakers, the duration of the post-atrial pacing ventricular safety window is 100 ms

In this tracing we observe post-VP and post-VS T wave oversensing and AV crosstalk with pacing at the end of the safety window (100 ms)

Should have chosen
In Biotronik pacemakers, the duration of the post-atrial pacing ventricular safety window is 110 ms
Selected
In Biotronik pacemakers, the duration of the post-atrial pacing ventricular safety window is 100 ms

GOOD ANSWERS :

  • Cross-talk AV
  • T-wave oversensing
  • In Biotronik pacemakers, the duration of the post-atrial pacing ventricular safety window is 100 ms

 

In this tracing we observe post-VP and post-VS T wave oversensing and AV crosstalk with pacing at the end of the safety window (100 ms)

Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
There is a p wave oversensing
The atrial sensitivity is programmed (value) too high
Selected
There is a R wave double counting

GOOD ANSWERS :

  • There is a R wave double counting
  • There is a VT

 

•There is a double counting of the R wave during a VT episode (V>A) that must be corrected by increasing the RV post sense blanking. The classification is appropriated despite the double counting.
•First and third A events are masked by the Far‑field  protection after Vs. 
Should have chosen
RV Blanking after atrial sensing must be increased
Selected
There is a VT

GOOD ANSWERS :

  • There is a R wave double counting
  • There is a VT

 

•There is a double counting of the R wave during a VT episode (V>A) that must be corrected by increasing the RV post sense blanking. The classification is appropriated despite the double counting.
•First and third A events are masked by the Far‑field  protection after Vs. 
Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
Pacing mode is DDDR

GOOD ANSWERS :

  • Pacing mode is DDDR
  • In this tracing, V-Epsd (blue star) indicates that the VT detection window is satisfied (8/10)

 

•Pacing mode is DDDR (see AP-Sr marker indicating Atrial pacing at sensor rate)
•The 7th marker As (red star) is sensed in the PVARP because of a Wenckebach behavior of device (spontaneous atrial rate too fast)
•There is no post A pacing ventricular safety window (noise window instead)
•V-Epsd indicates that the VT (not VF) detection window is satisfied (8/10)
Should have chosen
Selected
The 7th marker As (red star) is sensed in the PVARP, which has been extended to 500ms by the anti-PMT algorithm
Selected
In Boston ICDs, there is a post A pacing ventricular safety window
Selected
In this tracing, V-Epsd (blue star) indicates that the VF detection window is satisfied (8/10)
In this tracing, V-Epsd (blue star) indicates that the VT detection window is satisfied (8/10)

GOOD ANSWERS :

  • Pacing mode is DDDR
  • In this tracing, V-Epsd (blue star) indicates that the VT detection window is satisfied (8/10)

 

•Pacing mode is DDDR (see AP-Sr marker indicating Atrial pacing at sensor rate)
•The 7th marker As (red star) is sensed in the PVARP because of a Wenckebach behavior of device (spontaneous atrial rate too fast)
•There is no post A pacing ventricular safety window (noise window instead)
•V-Epsd indicates that the VT (not VF) detection window is satisfied (8/10)
Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
The pacing mode is VDI during the duration

GOOD ANSWERS : 

  • The pacing mode is VDI during the duration
  • V-Dur indicates the end of the VT duration
  • Quick Convert™ is engaged after V-Detect

 

•The ICD switches to VDI after the V-Epsd marker (not programmable)
•The end of VT duration (V-Dur marker) occurs before the end of the VF duration (V-Detect)
•VF episodes do not require A/V rate analysis (V rate only)
•Quick Convert™ is engaged after V-Detect and before charging (charge starts 300ms before the « Chrg » marker)
 
Should have chosen
V-Dur indicates the end of the VT duration

GOOD ANSWERS : 

  • The pacing mode is VDI during the duration
  • V-Dur indicates the end of the VT duration
  • Quick Convert™ is engaged after V-Detect

 

•The ICD switches to VDI after the V-Epsd marker (not programmable)
•The end of VT duration (V-Dur marker) occurs before the end of the VF duration (V-Detect)
•VF episodes do not require A/V rate analysis (V rate only)
•Quick Convert™ is engaged after V-Detect and before charging (charge starts 300ms before the « Chrg » marker)
 
Should have chosen
Selected
a VF episode is detected because V>A
Selected
Quick Convert™ is engaged after V-Detect

GOOD ANSWERS : 

  • The pacing mode is VDI during the duration
  • V-Dur indicates the end of the VT duration
  • Quick Convert™ is engaged after V-Detect

 

•The ICD switches to VDI after the V-Epsd marker (not programmable)
•The end of VT duration (V-Dur marker) occurs before the end of the VF duration (V-Detect)
•VF episodes do not require A/V rate analysis (V rate only)
•Quick Convert™ is engaged after V-Detect and before charging (charge starts 300ms before the « Chrg » marker)
 
Should have chosen
Selected
The ICD begins the charge during the Quick Convert™ delivery
Question

Your answerChoiceFeedbackCorrect answer
Secure Sense is triggered by 2 out of 3 fast ventricular events

GOOD ANSWER : Secure Sense is triggered by 2 out of 3 fast ventricular events

 

•There is a crosstalk (sensing of the atrial output in the ventricular channel resulting in inhibition of the ventricular pulse)
•Lead pacing/sensing configurations are always bipolar on ICDs
•The ventricular safety window is indeed 64ms, but cannot be extended
Should have chosen
Selected
There is a T wave oversensing
The A pacing configuration is probably Unipolar
There is a ventricular bigeminy
The ventricular safety window of 64ms can be extended
Question

Your answerChoiceFeedbackCorrect answer
Selected
BV and VS intervals reset the VT counter

GOOD ANSWER :

  • BV and VS intervals reset the VT counter
  • Change from true bipolar to integrated bipolar sensing may solve the problem

 

•The T-wave oversensing algorithm is applied after VT or VF has been detected to avoid inappropriate VT/VF therapy. It does not apply during consistent ventricular pacingCRT is effective every 2 beats
•Increasing the decay delay may prevent this oversensing in St Jude Medical devices (not available in Medtronic ICDs)
•Change from true bipolar to integrated bipolar sensing may solve the problem. The change from true bipolar to integrated bipolar sensing, permitted in new Medtronic devices, can sometimes lead to an increase of the R-wave amplitude and an increase in the ratio R/T wave amplitude that solve the problem.
Should have chosen
Selected
Turning on the T wave discrimination algorithm will eliminate this oversensing phenomenon
Selected
CRT is effective every 3 beats
Selected
To prevent this oversensing in this ICD, one can increase the decay delay
Selected
Change from true bipolar to integrated bipolar sensing may solve the problem

GOOD ANSWER :

  • BV and VS intervals reset the VT counter
  • Change from true bipolar to integrated bipolar sensing may solve the problem

 

•The T-wave oversensing algorithm is applied after VT or VF has been detected to avoid inappropriate VT/VF therapy. It does not apply during consistent ventricular pacingCRT is effective every 2 beats
•Increasing the decay delay may prevent this oversensing in St Jude Medical devices (not available in Medtronic ICDs)
•Change from true bipolar to integrated bipolar sensing may solve the problem. The change from true bipolar to integrated bipolar sensing, permitted in new Medtronic devices, can sometimes lead to an increase of the R-wave amplitude and an increase in the ratio R/T wave amplitude that solve the problem.
Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
Pacing mode is possibly DDI
The tracing is compatible with a normal functioning of the Saint Jude Medical Ventricular Intrinsic Preference (VIP™) algorithm
Selected
If the device would have been a LivaNova pacemaker programed with AAISafeR ™: the pacing mode would probably have switched to DDD

GOOD ANSWERS :

  • If the device would have been a LivaNova pacemaker programed with AAISafeR ™: the pacing mode would probably have switched to DDD
  • The maximal Ventricular sensitivity is 2mV when programmed to “AUTO”, in this Biotronik PM

 

•In DDI mode, each AP is followed by either a sensed or a paced V. There is a Ventricular undersensing.
•The VIP doesn’t accept blocked Ap. It is a AVD hysteresis.
•AAISafeR switch from ADI to DDD when 3 P waves are not followed by a sensed Ventricle in a 12 A intervals window
•MVP from ADI to DDD when 2 P waves are not followed by a sensed Ventricle in a 4 A intervals window
•The maximal Ventricular sensitivity is 2mV when programmed to “AUTO”, in this Biotronik PM
Should have chosen
If the device would have been a Medtronic pacemaker programed with MVP ™: the pacing mode would probably have switched to DDD
Selected
The maximal Ventricular sensitivity is 2mV when programmed to “AUTO”, in this Biotronik PM

GOOD ANSWERS :

  • If the device would have been a LivaNova pacemaker programed with AAISafeR ™: the pacing mode would probably have switched to DDD
  • The maximal Ventricular sensitivity is 2mV when programmed to “AUTO”, in this Biotronik PM

 

•In DDI mode, each AP is followed by either a sensed or a paced V. There is a Ventricular undersensing.
•The VIP doesn’t accept blocked Ap. It is a AVD hysteresis.
•AAISafeR switch from ADI to DDD when 3 P waves are not followed by a sensed Ventricle in a 12 A intervals window
•MVP from ADI to DDD when 2 P waves are not followed by a sensed Ventricle in a 4 A intervals window
•The maximal Ventricular sensitivity is 2mV when programmed to “AUTO”, in this Biotronik PM
Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
There is a intermitent ventricular loss of capture
Selected
You propose to increase the max tracking rate

GOOD ANSWERS :

  • You propose to increase the max tracking rate
  • You propose to increase the A-tachy response (ATR) trigger rate
  • You propose to lower the value of the maximal V sensitivity

 

•There is no loss of V capture. The last Vp beat (VP-FB) is not effective because delivered just after an undersensed intrinsic QRS.
•The MTR should be increased to have a 1:1 tracking of the sinus tachycardia
•The ATR trigger rate should be increased to avoid inapproriate fallback.
•Entry count of ATR is 8 max.
•There is a V undersensing which must be fixed by decreasing the value of the V sensitivity.
Should have chosen
You propose to increase the A-tachy response (ATR) trigger rate

GOOD ANSWERS :

  • You propose to increase the max tracking rate
  • You propose to increase the A-tachy response (ATR) trigger rate
  • You propose to lower the value of the maximal V sensitivity

 

•There is no loss of V capture. The last Vp beat (VP-FB) is not effective because delivered just after an undersensed intrinsic QRS.
•The MTR should be increased to have a 1:1 tracking of the sinus tachycardia
•The ATR trigger rate should be increased to avoid inapproriate fallback.
•Entry count of ATR is 8 max.
•There is a V undersensing which must be fixed by decreasing the value of the V sensitivity.
Should have chosen
Selected
You propose to increase the entry count of the ATR 
You propose to lower the value of the maximal V sensitivity

GOOD ANSWERS :

  • You propose to increase the max tracking rate
  • You propose to increase the A-tachy response (ATR) trigger rate
  • You propose to lower the value of the maximal V sensitivity

 

•There is no loss of V capture. The last Vp beat (VP-FB) is not effective because delivered just after an undersensed intrinsic QRS.
•The MTR should be increased to have a 1:1 tracking of the sinus tachycardia
•The ATR trigger rate should be increased to avoid inapproriate fallback.
•Entry count of ATR is 8 max.
•There is a V undersensing which must be fixed by decreasing the value of the V sensitivity.
Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
Selected
Pacing mode is VP-suppression
Selected
You propose to switch ON the Enhanced T-Wave Suppression algorithm
You propose to increase the atrial sensitivity (lower value)
Selected
The device has potentially self induced the arrhythmia

GOOD ANSWERS :

  • The device has potentially self induced the arrhythmia
  • ATP one shot is not delivered because the VF is unstable

 

•Pacing mode is DDI. There is a R wave undersensing with a Vp on the T wave. Atrial signal are well detected unless in the PVAB.
•The ryhthm in the VF zone must be regular (12%) for the ATP one shot to be delivered. Here stability is 51ms>12% of 251ms (30ms).
Should have chosen
Selected
ATP one shot is not delivered because the VF is unstable

GOOD ANSWERS :

  • The device has potentially self induced the arrhythmia
  • ATP one shot is not delivered because the VF is unstable

 

•Pacing mode is DDI. There is a R wave undersensing with a Vp on the T wave. Atrial signal are well detected unless in the PVAB.
•The ryhthm in the VF zone must be regular (12%) for the ATP one shot to be delivered. Here stability is 51ms>12% of 251ms (30ms).
Should have chosen
Question

Your answerChoiceFeedbackCorrect answer
Selected
The most likely diagnosis is a RV lead fracture
DDI is the « episode » pacing mode

GOOD ANSWER : DDI is the « episode » pacing mode

 

•Noise is present on 3 different channels which renders the diagnosis of lead fracture unlikely. This tracing is typical of electrocautery interference. The device switches from DDD to DDI after 3 fast intervals. Return to sinus is programmed to 3. ATP during charge is not delivered because the RR intervals are too short (<240ms). The discrimination channel is also contaminated, therefore the Secure Sense counter is reseted every 2 fast cycles.
Should have chosen
Return of sinus rhythm counter is programmed to 5 (nominal)
ATP during charge is not delivered because the VF is unstable
Selected
There is a rupture of the coil conductor
Question

Your answerChoiceFeedbackCorrect answer
ODO 
AAI/VVI Rythmiq 75bpm
Selected
AAI 75bpm

GOOD ANSWER : AAI 75bpm

Atrial pacing without V synchronization. Rhytmiq would have triggered ventricular pacing at 1000ms interval (602/min).

Should have chosen
DDI 75 bpm
DDD 75 bpm
Question

Your answerChoiceFeedbackCorrect answer
Selected
The ICD uses the combined counter to diagnose a VF episode (page 1)
The Fast VT zone is programmed via VF

GOOD ANSWERS :

  • The Fast VT zone is programmed via VF
  • First shock is not delivered because the device failed to reconfirm the arrhytmia after the end of the charge
  • In this case after VF redetection, the shock is commited

 

•The VF is classified using the FVT counter, not the combined counter. At the end of the FVT NID (=VF NID), the device reviews the last 8 V intervals, looking for FS intervals: if even one of the 8 is FS, the device classifies the episode as VF (if not, it classifies as FVT).
•FVT is programmed via VF. See the « TF∙ »markers. When programmed via VT, the marker is « T∙F ». Morover, the first classification is FD, which would not be possible with a « FVT via VT » programmation.
•After CE (charge end) the device is looking for two fast events out of five. There are 5 V intervals before « aborted »: two have to be shorter than the lowest VT zone + 60ms. Since the longest of these four is 520ms, then the lowest bound of the VT zone is <460ms.
•On the Virtuoso platform, there is no confirmation of the shock after a first abort.
•The shock is synchronized on a R wave.
Should have chosen
Selected
First shock is not delivered because the device failed to reconfirm the arrhytmia after the end of the charge

GOOD ANSWERS :

  • The Fast VT zone is programmed via VF
  • First shock is not delivered because the device failed to reconfirm the arrhytmia after the end of the charge
  • In this case after VF redetection, the shock is commited

 

•The VF is classified using the FVT counter, not the combined counter. At the end of the FVT NID (=VF NID), the device reviews the last 8 V intervals, looking for FS intervals: if even one of the 8 is FS, the device classifies the episode as VF (if not, it classifies as FVT).
•FVT is programmed via VF. See the « TF∙ »markers. When programmed via VT, the marker is « T∙F ». Morover, the first classification is FD, which would not be possible with a « FVT via VT » programmation.
•After CE (charge end) the device is looking for two fast events out of five. There are 5 V intervals before « aborted »: two have to be shorter than the lowest VT zone + 60ms. Since the longest of these four is 520ms, then the lowest bound of the VT zone is <460ms.
•On the Virtuoso platform, there is no confirmation of the shock after a first abort.
•The shock is synchronized on a R wave.
Should have chosen
Selected
In this case after VF redetection, the shock is commited

GOOD ANSWERS :

  • The Fast VT zone is programmed via VF
  • First shock is not delivered because the device failed to reconfirm the arrhytmia after the end of the charge
  • In this case after VF redetection, the shock is commited

 

•The VF is classified using the FVT counter, not the combined counter. At the end of the FVT NID (=VF NID), the device reviews the last 8 V intervals, looking for FS intervals: if even one of the 8 is FS, the device classifies the episode as VF (if not, it classifies as FVT).
•FVT is programmed via VF. See the « TF∙ »markers. When programmed via VT, the marker is « T∙F ». Morover, the first classification is FD, which would not be possible with a « FVT via VT » programmation.
•After CE (charge end) the device is looking for two fast events out of five. There are 5 V intervals before « aborted »: two have to be shorter than the lowest VT zone + 60ms. Since the longest of these four is 520ms, then the lowest bound of the VT zone is <460ms.
•On the Virtuoso platform, there is no confirmation of the shock after a first abort.
•The shock is synchronized on a R wave.
Should have chosen
The def shock is synchronized on a p wave