Chapter 4 · Oversensing · Case 7
Oversensing from the use of a electrocautery knife
Another form of interference is the current induced by a diathermy knife in the vicinity or not of a defibrillator. The current is detected on all channels, the signals are chaotic, and of variable amplitude.
The same risks are to be feared (inappropriate shock, and inhibition of pacing in device-dependent patients). However, the solution is extremely simple: during the use of the diathermy knife, just apply a magnet next to the defibrillator case which will inhibit both the detection of arrhythmias as well as therapies.
The pacing function is not changed on the latest MicroPort models (Platinium). On previous platforms, pacing was delivered in DDD or VVI mode according to the permanent pacing mode programmed at the magnet rate depending on the state of wear of the battery (96/min at the beginning of life (Beginning Of Service) and 8O/min at the time of replacement (Recommended Replacement Time)). In contrast, for pacemaker-dependent patients only, there is a function that allows pacing when noise (electrical interference with a rate greater than 16 Hz) is detected and properly managed, i.e. masked by the automatic retriggering of refractory periods (30 ms). At the end of the escape interval, if the refractory periods continue to be retriggered, pacing is delivered. This corresponds to functioning in asynchronous mode (risk-free for a pacemaker-dependent patient). The programmed mode is restored as soon as the interference is no longer detected. This function is accessible from the “Settings” screen, “Brady” tab, in the “Advanced Settings” section. After clicking on “Functioning during noise”, it will be necessary to activate the “V pacing on noise”.


Take-home message
- To avoid the harmful effect of a diathermy knife on the defibrillator, you must apply a magnet next to the device.
- To regain all functions, simply remove the magnet.
- In pacing-dependent patients, program the “V Pacing on noise” feature to avoid inhibition during noise
- The patient must be informed.
Practice out loud
Just describe the tracing. What do you think is going on? Describing the numbers can help.
—
From Implantable Cardioverter Defibrillator — Clinical cases from patients with MicroPort™ ICDs by M. Strik, A. Thiyagarajah, S. Ploux, P. Bordachar, P. Ritter. Published by Cardiocases. Each case is a question on real device interrogations: pick your answer, check it, then read the authors’ interpretation. Tap any figure to open it full screen.