              Same patient as previous tracing; device programmed to DDD mode 70 bpm;
          Trace
              Atrial unipolar pacing (high-amplitude stimulus) and ventricular unipolar pacing after a fixed atrioventricular delay;
          Trace
              Repeated AP-VP cycles (atrial and ventricular pacing) at 70 bpm;
          Exergue
              DDD mode is designed to respond to the characteristics presented by all implanted patients as a whole but may be associated with a high percentage of deleterious ventricular pacing in patients with preserved atrioventricular conduction.
           
 
 
 
The DDD mode is the standard programming mode of dual-chamber pacemakers or resynchronization devices. It ensures atrioventricular synchronization at rest as well as during exercise during sensed or paced atrial activity.