Dual chamber discrimination: V>A

Tracing
N° 46
Manufacturer Abbott Device ICD Field Discrimination
Patient

This 77-year-old man, recipient of a St Jude medical CRT-D implanted for idiopathic dilated cardiomyopathy, a LVEF at 30%, a wide QRS and a prolonged PR, was seen for a routine follow-up. 

Main programmed parameters

  • VF zone at 231 bpm, VT-2 zone at 181 bpm, a VT-1 zone at 160 bpm
  • 12 cycles in the VF zone, 12 cycles in the VT-2 zone and 12 cycles in the VT-1 zone are needed for the diagnosis
  • Effective discrimination in the VT-1 and VT-2 zones (dual chamber discrimination)
  • V<A: if one of the criteria is fulfilled; morphology (60%, 5 out of 8), stability (50 ms, with 60 ms delta AV association, 12 intervals)
  • V=A: if all criteria are fulfilled; morphology (60%, 5 out of 8), sudden onset (20%)
Graph and trace

Narrative

Episode of VT-1 in the V>A arm; no additional analysis and diagnosis of VT-1; ramp of 12 stimuli; probable success with slowing of the rate (740 ms coupling interval).

Tracing

  1. atrial and biventricular (AP-BP) pacing at sensor-indicated rate (SIR);
  2. tachycardia starts suddenly with atrioventricular dissociation; first cycles unclassified;
  3. DDI mode episode after 3 T1 classified cycles; it is noteworthy that the ventricular morphology is strictly identical to that of the reference QRS (v, 100);
  4. diagnosis of VT in the V>A arm; ventricular rate faster than the atrial rate; no additional discrimination needed for the diagnosis of VT;
  5. ramp of 12 stimuli;
  6. successful ramp and diagnosis of return to sinus rhythm.
Comments

This tracing illustrates the likelihood of a conflict between single chamber and dual chamber discrimination parameters. The diagnosis of VT is clearly apparent from the presence of AV dissociation; the episode was accurately identified in the V>A arm. In contrast, the morphology criterion was misleading (100% similarity with the reference complex). In this patient, who was continuously biventricular paced, the device probably chose an ectopic complex as its reference.

 

Other articles that may be of interest to you
Dual-chamber DDI mode
EGM recordings