A. AMS occurs when the filtered atrial rate interval (FARI) < the atrial tachycardia detection interval
B. if actual PP ≤ FARI: FARI decremented by 38 ms
C. if actual PP> FARI: FARI incremented by 25 ms
D. return to synchronous mode when FARI > the maximum synchronous rate interval
E. return to synchronous mode when FARI > the atrial tachycardia detection interval
A. this is an atrial tachycardia episode
B. this is a VT episode
C. this is a junctional tachycardia episode
D. this is a PMT episode
E. this is a fallback failure due to atrial arrhythmia
A. the AMS occurs in the setting of atrial arrhythmia
B. the AMS occurs in the setting of crosstalk
C. the AMS occurs in the setting of the oversensing of a 50 Hz signal
D. the AMS occurs in the setting of the oversensing of noise on the atrial lead
E. the AMS occurs in the setting of sinus tachycardia