post-ventricular atrial blanking and 2:1 flutter
Tracing
Manufacturer Abbott
Device PM
Field DAV, TRE, périodes réfractaires
N° 27
![](https://www.cardiocases.com/sites/default/files/styles/trace_thumb_middle/public/pdfpreview/6834-4.periode_refractaire_trace_27.jpg?itok=Uwjki7r3)
Graph and trace
Continuation of the tachycardia; modification of the programming with shortening of the post-ventricular atrial blanking (vertical line); the second atrial signal is now sensed in the PVARP (AR); the filtered atrial interval quickly reaches the switching threshold; DDIR mode and progressive deceleration of the ventricular pacing rate toward the sensor rate.