Unique and guided training + 800 clinical cases in the data bank
A. the programmed mode is VVI 60 beats/minute
B. the programmed mode is VVI 70 beats/minute
C. there is a failure of right ventricular capture
D. there is a failure of ventricular sensing
E. the intrinsic rhythm of the patient is a complete atrioventricular block
A. this is a manual ventricular threshold measurement by the operator
B. this is an automatic ventricular threshold measurement by the device
C. the ventricular threshold measurement is based on the analysis of the evoked response
D. the threshold value is measured at 0.5 V
E. the threshold value is measured at 0.625 V
A. in the most recent pacemakers, ventricular AutoCaptureTM is only programmable with a bipolar lead
B. the ventricular pacing threshold is automatically measured every 20 hours
C. in the event of loss of capture, a 5 V backup pulse is delivered 80 to 100 ms after the first stimulus
D. the ventricular pacing amplitude is automatically adjusted with a programmable safety margin between 0.25 V and 1.5 V
E. the pacemaker automatically adjusts the delivered amplitude and pulse duration
A. there is a failure of atrial sensing
B. there is a failure of atrial capture
C. there is a retrograde atrial conduction
D. there is an onset of PMT
E. there is presence of crosstalk
A. capture control is based on the morphology analysis of the evoked atrial response
B. the maximum amplitude that can be delivered is 5 Volts
C. there is a cycle-to-cycle control of atrial capture all throughout the day
D. the safety margin varies according to the measured threshold
E. the safety margin is programmable
A. dislodgement of the atrial lead
B. dislodgement of the ventricular lead
C. inversion of the 2 leads
D. increase in atrial pacing threshold
E. increase in ventricular pacing threshold